BEKO home - BEKO world - BEKO extra - Imprint - Legal Notes - Print
Company - Contact - News & Infos - Products

   
Products
Products Downloads Condensate discharge Condensate processing Filtration Drying Process technology Distribution Measurement Oil vapour Dewpoint Function Technical Data DP 207 Technical Data DP 109 Delivery Scope DP 207 Flow rate Leakages Min. pressure valve Service Training Tools
 
 

Mobile dew point measuring

DP 207 - Technical data
Measuring ranges -80... 50°C td, -20... 70°C, 0... 100 % RH
further parameters can be programmed by BEKO TECHNOLOGIES upon customer request:
ppmV/V, atmospheric dew point, g/m³, mg/m³, g/kg, or °F
Accuracy ± 0.5°C td (-10... 50°C td), typical ± 2 °C td at -40°C td
Pressure range Mounting without measuring chamber: -1 to 50 bar
Mounting with measuring chamber: -1 to 16 bar
Operating temperature -20... 70°C measuring gas temperature,
0... 50°C ambient temperature
Measuring chamber connection Internal thread G1/2“ / G1/4“,
Plug nipple NW 7,2; G1/4“ external,
for quick-release coupling
Screw-in thread R 1/2" stainless steel
Purge air demand 16 l/min when the measuring chamber is installed
(for 7 bar)
Housing material Polycarbonate
Protection class IP 65
EMV DIN EN 61326
Weight 340 g (without measuring chamber)
Power supply Internal NiMH battery for approx. 15 h operation, indication "Low Bat. max 2 h”, residual operation time, charging time: max. 2 h
Power supply unit 100 – 240 VAC – 50/ 60 Hz, 0.2 A,
24 VDC – 300 mA
Safety class 2, only for application in dry rooms
Display Single-line, resolution 0.1
Dew point (°C td), Relative humidity (% RH),
Temperature (°C or °F) according to factory setting,
Max. 3 parameters can be chosen one after the other by means of the scroll function "arrow down"
   
Display functions Min., Max., battery status
Field of application Designed for measurement downstream of a compressed air dryer.